MEASUREMENTS OF ELONGATIONAL VISCOMETRY USING A FIBER SPINNING TECHNIQUE. PART I. MODIFICATIONS TO THE SANGAMO SCHLUMBERGER ELONGATIONAL VISCOMETER MODEL E4

Authors
  1. Mayer, M.D.G.
  2. Fenrick, W.J.
  3. Armour, S.J.
Corporate Authors
Defence Research Establishment Suffield, Ralston ALTA (CAN)
Abstract
The Sangamo-Schlumberger Elongational Viscometer (Model E4), as delivered by the manufacturer, did not maintain a constant elongational load under constant flow conditions and consequently could not be used to accurately measure elongational viscosity. The extensive modifications made to the instrument at DRES to correct the problem are described in detail as are other improvements made to the instrument.
Report Number
377 —
Date of publication
15 Feb 1984
Number of Pages
44
DSTKIM No
84-01785
CANDIS No
121182