A RECORDING IONIZATION CHAMBER INSTRUMENT

Authors
  1. Cunningham, J.R.
  2. Wilson, R.
Corporate Authors
Defence Research Chemical Labs, Ottawa ONT (CAN)
Abstract
A sensitive recording ionization chamber with three ranges, 0 - 1, 0 - 10, and 0 - 100 mr/hr, is described. The d.c. amplifier uses 100% negative feedback to achieve good stability and linearity. The high sensitivity is achieved partly through use of a large volume (5.5 litres) cylindrical chamber. The presence of a dip in the polar response curve in the direction along the axis of the chamber is discussed.
Report Number
DRCL-261 — Previously accessioned as DSIS 58/10452
Date of publication
15 May 1958
Number of Pages
14
DSTKIM No
92-03800
CANDIS No
127146
Format(s):
Hardcopy;Originator's fiche received by DSIS