IMPROVEMENTS TO THE RADIOMETRIC CALIBRATION OF THE COMPACT AIRBORNE SPECTROGRAPHIC IMAGER

Authors
  1. Babey, S.K.
  2. Soffer, R.J.
Corporate Authors
Defence Research Establishment Suffield, Ralston ALTA (CAN);ITRES Research Ltd, Calgary ALTA (CAN)
Abstract
The CASI is an airborne pushbroom imaging spectrograph sensitive in the visible and near infrared portion of the spectum. The ability to reliably obtain calibrated radiances at high spectral and spatial resolution is crucial to many of the imaging spectroscopy studies that are being carried out with this instrument. This report describes a new calibration procedure that has been developed in order to improve the accuracy with which the casi can be calibrated while at the sme time simplifing the approach. This simplification of the algorithms has allowed a more detailed investigation of the individual steps of the calibration to be carried out, the results of which are presented here. The accuracy and signal to noise levels resulting from use of the new procedures are also assessed.
Keywords
CASI - COMPACT AIRBORNE SPECTROGRAPHIC IMAGER
Report Number
DRES-CR-35-92 — Contract Report (Final)