CALIBRATION OF ACCELEROMETERS USING THE RATE TABLE AT THE DCIEM IMPACT STUDIES FACILITY

Authors
  1. Bowden, T.
Corporate Authors
Defence and Civil Inst of Environmental Medicine, Downsview ONT (CAN)
Abstract
A method for calibrating linear accelerometers using the rate table at the DCIEM Impact Studies Facility (ISF) computer is described. The method uses the Datalab transient recorders at the ISF to perform up to 4096 measurements of the accelerometer output voltage and the rate table angular velocity, as the angular velocity is varied. The data from these measurements are used to calculate the slope and intercept of the line of best fit relating the measured output voltage to the centripetal acceleration calculated from the angular velocity. A computer program controls the process of data transfer, data selection and data reduction. The accelerormeter output voltage can be plotted as a function of acceleration, if required.
Report Number
85-C-14 — Technical Communication
Date of publication
15 May 1985
Number of Pages
24
DSTKIM No
85-02182
CANDIS No
46984